FDA Recall Terminated

Fresenius Liberty Cycler Part Number: RTLR180111 Product Usage: The Liberty cycler is a portable device designed to perform acute and chronic peritoneal dialysis.

Recall: Z-0148-2014 · Initiated October 14, 2013

Recall

Recall Number
Z-0148-2014
Event Number
66339
Firm
Fresenius Medical Care Holdings, Inc.
FEI Number
3001451489
Product Code
FKX
Status
Terminated
Root Cause
Device Design
Initiated
October 14, 2013
Posted
November 5, 2013
Terminated
May 18, 2021
Address
920 Winter St, Waltham, MA, 02451-1521

Description

Fresenius Liberty Cycler Part Number: RTLR180111 Product Usage: The Liberty cycler is a portable device designed to perform acute and chronic peritoneal dialysis.

Reason

Fluid leaking may be observed inside the pump compartment of the Liberty Cycler and may result in dialysate becoming contaminated, potentially causing peritonitis

Action

Fresenius Medical Care contacted all affected consignees by telephone on 10/14/13 to inform them of the product safety alert. Follow-up Urgent Safety Alert letter was sent on 10/18/13 and 10/23/13 by Certified Mail, Return Receipt requested. The letter identified the affected product, problem and actions to be taken. Customers were instructed to report any events associated with this safety alert to Customer Service Team at 1-800-323-5188.

Distribution

USA Nationwide Distribution

Quantity

16,600