FDA Recall
Terminated
AXIOM Artis Systems VB23A-VB23H with Perivision functionality, Model Numbers: 5904433,5904649, 5904466, 5904441, 5904656, 7728350, and 7007755 Angiographic X-ray System
Recall: Z-0144-2011
·
Initiated September 9, 2010
Recall
- Recall Number
- Z-0144-2011
- Event Number
- 56754
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- IZI
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- September 9, 2010
- Posted
- October 26, 2010
- Terminated
- September 29, 2012
- Address
- 51 Valley Stream Pkwy, Malvern, PA, 19355
Description
AXIOM Artis Systems VB23A-VB23H with Perivision functionality, Model Numbers: 5904433,5904649, 5904466, 5904441, 5904656, 7728350, and 7007755 Angiographic X-ray System
Reason
Sporadic system crash while Perivision is performed.
Action
Siemens issued an Urgent Field Safety Notice to all customers. The letter informed customers of the potential issues and provided additional information on the implemented resolution. To resolve the problem, a system update was to be installed on each affected system under Update Instructions AX054/10/S. For questions phone (610) 448-3237.
Distribution
USA - Nationwide Distribution
Quantity
286 units