FDA Recall Terminated

Merge Healthcare, Merge Hemo Programmable diagnostic computer. The system comprises the Patient Data Module and the HeartSuite Hemodynamics Hemo Monitor PC. The two units are connected via a serial interface. All vital parameters and evaluations are registered and calculated in the Patient Data Module. This data is then transmitted to the HeartSuite Hemodynamics Hemo Monitor PC via the serial interface. All data can be shown and monitored on the HeartSuite Hemodynamics Hemo Monitor PC. The Patient Data Module uses an internal battery charged from an external power input (RS 232/12V). The power supply, like the data transmission, is completely isolated from the visualization unit. The HeartSuite Hemodynamics Hemo Monitor PC is powered via the normal mains connection 230V/1 IOV.

Recall: Z-0143-2015 · Initiated September 29, 2014

Recall

Recall Number
Z-0143-2015
Event Number
69406
Firm
Merge Healthcare, Inc.
FEI Number
1000116130
Product Code
DQK
Status
Terminated
Root Cause
Software design
Initiated
September 29, 2014
Posted
October 28, 2014
Terminated
August 25, 2016
Address
900 Walnut Ridge Dr, Hartland, WI, 53029-8347

Description

Merge Healthcare, Merge Hemo Programmable diagnostic computer. The system comprises the Patient Data Module and the HeartSuite Hemodynamics Hemo Monitor PC. The two units are connected via a serial interface. All vital parameters and evaluations are registered and calculated in the Patient Data Module. This data is then transmitted to the HeartSuite Hemodynamics Hemo Monitor PC via the serial interface. All data can be shown and monitored on the HeartSuite Hemodynamics Hemo Monitor PC. The Patient Data Module uses an internal battery charged from an external power input (RS 232/12V). The power supply, like the data transmission, is completely isolated from the visualization unit. The HeartSuite Hemodynamics Hemo Monitor PC is powered via the normal mains connection 230V/1 IOV.

Reason

It has been reported that during use, the SpO2 value displayed on the Hemo Monitor may not update to reflect changes in the patient's oxygen value. It is also possible that if using the pulse rate of the SpO2 finger clip to calculate the patient's heart rate, it too may be subject to displaying a stale value.

Action

Consignees were e mailed on 5/29/14 a Merge "Important Alert" letter. The letter described the problem and the product involved in the recall. Informed consignees that they had the opportunity to replace the Nellcor SpO2 module with a Masimo SpO2 module free of charge. The letter also provided Clinical Information and the Next Steps to follow. This included to contact Merge Technical Support if they wanted to replace their Nellcor module. For questions, they can contact Lee Harrop, Director Solutions Management, Cardiology, 386-239-2224, [email protected]. A second letter was sent Sept. 11, 2014 via e-mail. Consignees were made aware of a firmware patch available.

Distribution

Nationwide Distribution including the states of AZ, CA, FL, IL, IA, MA, MO, OH, SC, TN, and WI.

Quantity

52