FDA Recall Terminated

D-Stat Hemostat Band, model number 3500. Sterile, latex free single use device. Vascular Solutions, Inc. 6464 Sycamore Court, Minneapolis, MN 55369 USA. Each D-Stat Radial hemostatic band (D-Stat Radial) consists of the following components: -Lyophilized pad consisting of thrombin, sodium carboxylmethyulcellulose and calcium chloride in a nonwoven gauze; -Application device consisting of an adjustable retention strap, collar and attached gauze pad.

Recall: Z-0143-06 · Initiated September 9, 2005

Recall

Recall Number
Z-0143-06
Event Number
33456
Firm
Vascular Solutions, Inc.
FEI Number
3002827704
Product Code
FRO
Status
Terminated
Root Cause
Other
Initiated
September 9, 2005
Posted
November 16, 2005
Terminated
November 13, 2006
Address
6464 Sycamore Ct N, Osseo, MN, 55369-6032

Description

D-Stat Hemostat Band, model number 3500. Sterile, latex free single use device. Vascular Solutions, Inc. 6464 Sycamore Court, Minneapolis, MN 55369 USA. Each D-Stat Radial hemostatic band (D-Stat Radial) consists of the following components: -Lyophilized pad consisting of thrombin, sodium carboxylmethyulcellulose and calcium chloride in a nonwoven gauze; -Application device consisting of an adjustable retention strap, collar and attached gauze pad.

Reason

A defect affecting the seals of the paper pouch holding the adhesive bandage provided with the D-Stat products was noted during manufacturing. The integrity of the seals of the D-stat Dry pads which are packaged separately in a foil pouch is not in question.

Action

Urgent Medical Device Safety Information, dated 09/09/05, was sent to all affected customers via Fed Ex beginning 09/09/05. The letter informed the customers of the issue and askes customers to sequester affected product. A Vascular Solutions, Inc. Sales Representative will visit locations to remove and replace affected product beginning 09/19/05.

Distribution

Within the US to AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, PR, RI, SC, TX, UT, VA, WI and WV. OUS to include: Italy, United Kingdom

Quantity

415