FDA Recall Terminated

TITAN Operating Room Table Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.

Recall: Z-0141-2015 · Initiated October 9, 2014

Recall

Recall Number
Z-0141-2015
Event Number
69426
Firm
Trumpf Medical Systems, Inc.
FEI Number
3003184737
Product Code
JEA
Status
Terminated
Root Cause
Device Design
Initiated
October 9, 2014
Posted
October 27, 2014
Terminated
March 23, 2016
Address
1046 Legrand Blvd, Charleston, SC, 29492-7672

Description

TITAN Operating Room Table Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system.

Reason

Unintended movement of the operating table during surgical cases due to the use of damaged TITAN remotes.

Action

TRUMPF sent an Urgent Medical Device Correction letter dated October 3, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Actions to be taken: Complete and return the attached Confirmation of Receipt form with serial numbers of the TITAN table(s) and the serial number(s) for any remote controls you have. We will then arrange to send replacement remotes, which have been redesigned to eliminate this mode of failure. You will then be required to dispose of your recalled remotes. If you have any questions concerning this request or the procedure as outlined, please contact TRUMPF Medical Systems, Inc. at 888-474-9359.

Distribution

Worldwide Disttribution: US (nationwide) including the states of AL, AZ, CT, FL, GA, IA, IL, IN, KS, MD, MI, MN, MO, NC, ND, NJ, NM, NV, NY, OH, OR, PA, SC, TX, VA, WA and WY: and Internationally to Canada.

Quantity

184 units