Cardinal Health, Craig Calibrated Cannula, overall length 5-3/8" (13.7 cm), catalog #OS5000-005, sold separately, or as a component of the Craig Vertebral Biopsy Set, catalog #OS5000
Recall
- Recall Number
- Z-0141-2010
- Event Number
- 53137
- Firm
- Cardinal Health
- FEI Number
- 3001236905
- Product Code
- FCG
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- August 28, 2009
- Posted
- November 9, 2009
- Terminated
- May 12, 2010
- Address
- 1430 Waukegan Rd, Mc Gaw Park, IL, 60085-6726
Description
Cardinal Health, Craig Calibrated Cannula, overall length 5-3/8" (13.7 cm), catalog #OS5000-005, sold separately, or as a component of the Craig Vertebral Biopsy Set, catalog #OS5000
The Craig Calibrated Cannula does not interface with the toothed cutter properly, which may prevent the toothed cutter from properly advancing through the cannula to the biopsy site.
Cardinal Health sent Urgent Product Recall letters dated 8/28/09 to all customers who received the product during the recalled period on the same day via UPS Next Day Air. The accounts were informed of the problem with the cannula, and were instructed to functionally test the cannula by passing the toothed cutter within it to ensure the cutter advances freely into the cannula. If any restriction is noticed, the product should be segregated and returned to Cardinal Health. The accounts were requested to complete and return by fax the enclosed acknowledgement form, indicating the number of affected cannulas being returned. Any questions were directed to V. Mueller Customer Service at 800-323-9088.
Nationwide distribution.
19 units