Patient Cart used in conjunction with the da Vinci S surgical and SI System. Intuitive Surgical 1266 Kifer Road Sunnyvale, CA 94086 The Patient Side Cart (PSC) is a robotic platform used for da Vinci Surgery. It is the operative component for both da Vinci systems and its primary function is to support he instrument arms and camera arm. It uses remote center technology. The remote center is a fixed point in space around which the Patient Cart arms move. Remote center technology enables the system to maneuver instruments and endoscopes in the surgical site while exerting minimal force on the patient body wall. The Patient Cart works in the sterile field, assisting the Surgeon Console operator by exchanging instruments and endoscopes, and by performing other patient-side activities. To help ensure patient safety, the action of the Patient Cart operator take precedence over actions of the Surgeon Console operator.

Recall: Z-0139-2014 · Initiated November 30, 0012

Recall

Recall Number
Z-0139-2014
Event Number
66526
Firm
Intuitive Surgical, Inc.
FEI Number
3001675293
Product Code
NAY
Status
Terminated
Root Cause
Device Design
Initiated
November 30, 0012
Posted
November 4, 2013
Terminated
November 4, 2013
Address
1266 Kifer Rd, Bldg 100, Sunnyvale, CA, 94086-5304

Description

Patient Cart used in conjunction with the da Vinci S surgical and SI System. Intuitive Surgical 1266 Kifer Road Sunnyvale, CA 94086 The Patient Side Cart (PSC) is a robotic platform used for da Vinci Surgery. It is the operative component for both da Vinci systems and its primary function is to support he instrument arms and camera arm. It uses remote center technology. The remote center is a fixed point in space around which the Patient Cart arms move. Remote center technology enables the system to maneuver instruments and endoscopes in the surgical site while exerting minimal force on the patient body wall. The Patient Cart works in the sterile field, assisting the Surgeon Console operator by exchanging instruments and endoscopes, and by performing other patient-side activities. To help ensure patient safety, the action of the Patient Cart operator take precedence over actions of the Surgeon Console operator.

Reason

Increased number of complaints regarding fluid entering the patient card base of the da Vinci S and SI systems.

Action

Intuitive Surgical Inc sent an Product Notification to all affected customers via email on November 30, 2012. E-mail notifications advised users of an addendum to the User Manual for Covering the Patient Cart column that provided a recommended method for covering the da Vinci patient cart column to prevent splashing on the cart during a procedure. The addendum was attached to the e-mail and is available online. If you have any questions, please contact Customer Service at 1-800-876-1310. Outside the United States call +41 21 821 2020 or +800.0821.2020

Distribution

Worldwide Distribution - USA including Puerto Rico and the countries of Australia, Austria, Argentina, Belgium, Bulgaria, Brazil, Canada, Chile, China, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, India, Indonesia, Ireland, Israel, Japan, Kuwait, Mexico, Netherlands, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Uruguay and Venezuela.

Quantity

1754 patient carts