Dimension Vista V-LYTE Standard A, Catalog No K820 For the calibration of Na+/K+/Cl- on the Dimension vista System.
Recall
- Recall Number
- Z-0139-2011
- Event Number
- 56703
- Firm
- Siemens Healthcare Diagnostics, Inc.
- FEI Number
- 2517506
- Product Code
- CEM
- Status
- Terminated
- Root Cause
- Other
- Initiated
- August 19, 2010
- Posted
- October 25, 2010
- Terminated
- October 26, 2010
- Address
- 500 GBC Drive PO BOX 6101, Mailstop 514, Newark, DE, 19714-6101
Description
Dimension Vista V-LYTE Standard A, Catalog No K820 For the calibration of Na+/K+/Cl- on the Dimension vista System.
High potassium content in the affected lots results in the potential for low potassium recovery for Dilution Check, QC and patient samples when run on the Dimension Vista System.
Siemens sent an Urgent Field Safety Notice (dated "August 2010") to customers identifying the product, the problem and the action to be taken. Customers were instructed on how to continue testing until new lots of the recalled product were available and asked to complete and fax a Field Correction Effectiveness Check to 301-631-8467 to confirm receipt of the Urgent Field Safety Notice. For technical questions or concerns customers were instructed to contact the Siemens Technical Solutions Center at 800-441-9250.
USA, Nationwide Distribution
3,706 cartons