FDA Recall
Terminated
Stereotactic Circular Collimator - 3D Line stereotatic Hardware Accessories The device is part of the class of medical devices that are used in radiotherapy for the treatments of head tumors.
Recall: Z-0138-2014
·
Initiated July 23, 2013
Recall
- Recall Number
- Z-0138-2014
- Event Number
- 65782
- Firm
- Elekta, Inc.
- FEI Number
- 1037831
- Product Code
- HAW
- Status
- Terminated
- Root Cause
- No Marketing Application
- Initiated
- July 23, 2013
- Posted
- November 4, 2013
- Terminated
- November 18, 2014
- Address
- 400 Perimeter Center Ter Ne, Ste 50, Atlanta, GA, 30346-1227
Description
Stereotactic Circular Collimator - 3D Line stereotatic Hardware Accessories The device is part of the class of medical devices that are used in radiotherapy for the treatments of head tumors.
Reason
Potential for clinical errors.
Action
Elekta sent an Important Field Safety Notice dated 7/2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were advised of the issue and recommendation to resolve either by upgrading their system or to discontinue a clinical use of the device. For further questions please call ( 770 ) 300-9725.
Distribution
Nationwide distribution: USA including the states of AZ, CA, FL, GA, MO, NJ, NM, NC, OR, TN and WV.
Quantity
89