FDA Recall Terminated

Stereotactic Circular Collimator - 3D Line stereotatic Hardware Accessories The device is part of the class of medical devices that are used in radiotherapy for the treatments of head tumors.

Recall: Z-0138-2014 · Initiated July 23, 2013

Recall

Recall Number
Z-0138-2014
Event Number
65782
Firm
Elekta, Inc.
FEI Number
1037831
Product Code
HAW
Status
Terminated
Root Cause
No Marketing Application
Initiated
July 23, 2013
Posted
November 4, 2013
Terminated
November 18, 2014
Address
400 Perimeter Center Ter Ne, Ste 50, Atlanta, GA, 30346-1227

Description

Stereotactic Circular Collimator - 3D Line stereotatic Hardware Accessories The device is part of the class of medical devices that are used in radiotherapy for the treatments of head tumors.

Reason

Potential for clinical errors.

Action

Elekta sent an Important Field Safety Notice dated 7/2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were advised of the issue and recommendation to resolve either by upgrading their system or to discontinue a clinical use of the device. For further questions please call ( 770 ) 300-9725.

Distribution

Nationwide distribution: USA including the states of AZ, CA, FL, GA, MO, NJ, NM, NC, OR, TN and WV.

Quantity

89