Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System.
Recall
- Recall Number
- Z-0138-2010
- Event Number
- 52056
- Firm
- GE Medical Systems, LLC
- FEI Number
- 2126677
- Product Code
- JAA
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- April 16, 2009
- Posted
- November 9, 2009
- Terminated
- December 3, 2015
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System.
Potential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room monitor stating that three minutes are left before the X-ray is inhibited. This could lead to a delay in treatment and possible additional X-ray exposure and contra
Consignees were sent on May 1, 2009 a GE Healthcare " Urgent Medical Device Correction" letter dated April 16, 2009. The letter was addressed to Hospital Administrator/ Risk Manger, Mangers of Radiology/ Cardiology, and Radiologists/ Cardiologists. The letter described the problem, Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information.
Worldwide distribution: USA, ARGENTINA, BELGIUM, BRAZIL, BULGARIA, CHINA, CROATIA, EGYPT, FRANCE, GERMANY, HONG KONG, HUNGARY, INDIA, ITALY, JAPAN, REPUBLIC OF KOREA, LATVIA, LIBYA, MEXICO, NORWAY, POLAND, RUSSIA, SINGAPORE, SPAIN, SWEDEN, TAIWAN, TURKEY, UNITED KINGDOM, ROMANIA, ISRAEL, UKRAINE, YEMEN, BELARUS, and AUSTRALIA.
15 AND 46 UNLOCATED