Breast Biopsy Aiming Device. Magnetic Resonance Imaging System. Model number 5728735
Recall
- Recall Number
- Z-0138-06
- Event Number
- 33921
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2240869
- Product Code
- LNH
- Status
- Terminated
- Root Cause
- Other
- Initiated
- October 5, 2005
- Posted
- November 15, 2005
- Terminated
- August 7, 2007
- Address
- 51 Valley Stream Pkwy, Malvern, PA, 19355-1406
Description
Breast Biopsy Aiming Device. Magnetic Resonance Imaging System. Model number 5728735
If the recommended workflow is not followed and the worksheets are printed after moving the patient table out of isocenter, the offset values will be incorrect. This could result in a biopsy being performed at an incorrect location and presents the possibility of an inconclusive diagnosis. calculation of offset values for the breast biopsy aiming device may be incorrect
The recalling firm issued a Customer Safety Advisory via certified mail to affected customers per Update Instructions MR034/05/S. The letter re-emphasizes the importance of printing the offset values on the worksheet prior to moving the patient table and in accordance with the User's Manual.
The product was shipped to medical facilities in CA, FL, IL, MD, MN, NM, OH, OR, RI, and WI.
16 units