FDA Recall Terminated

Breast Biopsy Aiming Device. Magnetic Resonance Imaging System. Model number 5728735

Recall: Z-0138-06 · Initiated October 5, 2005

Recall

Recall Number
Z-0138-06
Event Number
33921
Firm
Siemens Medical Solutions USA, Inc
FEI Number
2240869
Product Code
LNH
Status
Terminated
Root Cause
Other
Initiated
October 5, 2005
Posted
November 15, 2005
Terminated
August 7, 2007
Address
51 Valley Stream Pkwy, Malvern, PA, 19355-1406

Description

Breast Biopsy Aiming Device. Magnetic Resonance Imaging System. Model number 5728735

Reason

If the recommended workflow is not followed and the worksheets are printed after moving the patient table out of isocenter, the offset values will be incorrect. This could result in a biopsy being performed at an incorrect location and presents the possibility of an inconclusive diagnosis. calculation of offset values for the breast biopsy aiming device may be incorrect

Action

The recalling firm issued a Customer Safety Advisory via certified mail to affected customers per Update Instructions MR034/05/S. The letter re-emphasizes the importance of printing the offset values on the worksheet prior to moving the patient table and in accordance with the User's Manual.

Distribution

The product was shipped to medical facilities in CA, FL, IL, MD, MN, NM, OH, OR, RI, and WI.

Quantity

16 units