FDA Recall Terminated

Infusion Pump I.V. Administration, specifically Plum Microbore Infusion Sets with Inline Filters labeled as PRIMARY PLUMSET 15 Micron Filter in Sight Chamber, CLAVE Port, 0.2 Micron Filter, CLAVE Y-Site, PE Lined Light Resistant Tubing, Distal Microbore Tubing 272 cm / 15 mL. 1 unit per pouch; 50 pouches per case. Not distributed in the U.S.

Recall: Z-0136-2020 · Initiated July 29, 2019

Recall

Recall Number
Z-0136-2020
Event Number
83549
Firm
ICU Medical Inc
FEI Number
3013319212
Product Code
FRN
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
July 29, 2019
Terminated
January 12, 2024
Address
600 N Field Dr, Lake Forest, IL, 60045-4835

Description

Infusion Pump I.V. Administration, specifically Plum Microbore Infusion Sets with Inline Filters labeled as PRIMARY PLUMSET 15 Micron Filter in Sight Chamber, CLAVE Port, 0.2 Micron Filter, CLAVE Y-Site, PE Lined Light Resistant Tubing, Distal Microbore Tubing 272 cm / 15 mL. 1 unit per pouch; 50 pouches per case. Not distributed in the U.S.

Reason

Potential for fluid leakage out of air vents on inline filters on Plum and Sapphire micropore infusion sets.

Action

ICU Medical notified customers, via Stericycle, Inc., on about 07/29/2019 via "URGENT: MEDICAL DEVICE CORRECTION" letter. Customers were instructed to check inventory and quarantine any sets that are affected by the recall, return them to ICU Medical, inform potential users within your organization of the recall, and complete and return the provided response form. Please notify customers if affected product was further distributed and request that they contact Stericycle at 1-855-311-5437 (M-F, 8am-5pm ET) to receive a reply form. Customers were also requested to consider the following situations: 1. In applications where the use of a filter is not deemed necessary, consider utilizing an administration set without a filter. 2. In applications where filter use is desired and microbore tubing is not required, consider using filtered macrobore sets in place of microbore sets. 3. In cases where a filtered macrobore set is desired but no alternative macrobore set is available contact ICU Medical because a non-filtered set along with an add on filter or filtered extension set could be considered. 4. In instances where the benefits of a filtered microbore infusion set are greater than the potential risks of a filter leak, the product affected by the recall may be utilized out of medical necessity. In these cases, an individualized assessment of benefits and risks should be undertaken with consideration of the unique characteristics of the patient and the infusion itself. When the benefits exceed the risks and the set is deemed medically necessary, users should periodically monitor the vented filter area for leakage. If leaking is observed, consider changing to a new infusion set as soon as clinically appropriate. Report this event to ICU Medical.

Distribution

Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY. International distribution to Canada, Chile, France, Germany, Greece, Hong Kong, Ireland, Italy, Oman, Philippines, Poland, Saudi Arabia, South Africa, Spain, Taiwan, Turkey, United Arab Emirates, United Kingdom.

Quantity

14,850 units