FDA Recall Terminated

AR ManoShield, MSS-3599, 10 pack. FGS: MSS-3599, Individual package label: LBL-110275-03, Box label: LBL-110282-03, Package Inserts: DOC-3682, User Manual: DOC-1332-12. Product Usage: The MVS disposable sheath (a.k.a. ManoShield) for the high-resolution MVS catheter is designed to provide a biocompatible barrier between the catheter and patient.

Recall: Z-0135-2012 · Initiated October 12, 2011

Recall

Recall Number
Z-0135-2012
Event Number
60175
Firm
Sierra Scientific Instruments Inc
FEI Number
3005344223
Product Code
FFX
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
October 12, 2011
Posted
October 28, 2011
Terminated
August 30, 2012
Address
5757 W Century Blvd, Ste 660, Los Angeles, CA, 90045-6456

Description

AR ManoShield, MSS-3599, 10 pack. FGS: MSS-3599, Individual package label: LBL-110275-03, Box label: LBL-110282-03, Package Inserts: DOC-3682, User Manual: DOC-1332-12. Product Usage: The MVS disposable sheath (a.k.a. ManoShield) for the high-resolution MVS catheter is designed to provide a biocompatible barrier between the catheter and patient.

Reason

The recall was initiated after Given Imaging received a test report during a Canadian licensing process where the AR ManoShield exhibited pinholes and did not pass the test, the Field Safety Corrective Action (FSCA) Committee was convened. As a result of the data discussed during this meeting, a decision was made to voluntarily recall this product, and cease its further commercialization or distr

Action

Given Imaging sent an "Urgent Product Recall 1st Notice" letter with an attached Device Recall Confirmation Form (via Fed Ex with signature for tracking purposes) to all customers who purchased the AR ManoShield, part numbers MSS-3599. The letter provides the customers with an explanation of the problem identified and actions to be taken. Customers were instructed to locate the affected product and removed from use immediately. The letter stated that if the affected product have been further distributed, forward this letter immediately to all customers who received the affected product. Customers were asked to complete and return the enclosed confirmation form as per the instructions. Customers with any questions were instructed to contact Karla Guerrero at (310) 641-8492 ext. 2138 or email at [email protected].

Distribution

Worldwide Distribution - USA (nationwide) and the countries of: Austria, Belgium, Canada, Chile, China, Columbia, Ecuador, France, Germany, Israel, Mexico, Poland, Portugal, Sweden, and Switzerland.

Quantity

292 units