GE Healthcare, GE Innova 3100 / 3100 IQ, Cardiovascular Imaging System, (The system consists of an a monoplane positioner, a vascular or cardiac table, an X-ray system and a digital detector)
Recall
- Recall Number
- Z-0135-2010
- Event Number
- 52056
- Firm
- GE Medical Systems, LLC
- FEI Number
- 2126677
- Product Code
- MQB
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- April 16, 2009
- Posted
- November 9, 2009
- Terminated
- December 3, 2015
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
GE Healthcare, GE Innova 3100 / 3100 IQ, Cardiovascular Imaging System, (The system consists of an a monoplane positioner, a vascular or cardiac table, an X-ray system and a digital detector)
Potential pump failure associated with the Coolix 4000 chiller of the Innova single plane and biplane systems that may stop the chiller from working. If a chiller failure occurs, the user will be informed with a message on the n-room monitor stating that three minutes are left before the X-ray is inhibited. This could lead to a delay in treatment and possible additional X-ray exposure and contra
Consignees were sent on May 1, 2009 a GE Healthcare " Urgent Medical Device Correction" letter dated April 16, 2009. The letter was addressed to Hospital Administrator/ Risk Manger, Mangers of Radiology/ Cardiology, and Radiologists/ Cardiologists. The letter described the problem, Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information.
Worldwide distribution: USA, ARGENTINA, BELGIUM, BRAZIL, BULGARIA, CHINA, CROATIA, EGYPT, FRANCE, GERMANY, HONG KONG, HUNGARY, INDIA, ITALY, JAPAN, REPUBLIC OF KOREA, LATVIA, LIBYA, MEXICO, NORWAY, POLAND, RUSSIA, SINGAPORE, SPAIN, SWEDEN, TAIWAN, TURKEY, UNITED KINGDOM, ROMANIA, ISRAEL, UKRAINE, YEMEN, BELARUS, and AUSTRALIA.