Siemens Dimension Vista Akaline Phosphatase (ALPI) Flex reagent cartridge - In vitro diagnostic test for the quantitative measurement of alkaline phosphatase in human serum and plasma Catalog Number: K2115, SMN# 10642444
Recall
- Recall Number
- Z-0134-2021
- Event Number
- 86430
- Firm
- Siemens Healthcare Diagnostics, Inc.
- FEI Number
- 2517506
- Product Code
- CJO
- Status
- Terminated
- Root Cause
- Under Investigation by firm
- Initiated
- August 25, 2020
- Posted
- October 20, 2020
- Terminated
- May 4, 2023
- Address
- 500 Gbc Dr, Newark, DE, 19702-2466
Description
Siemens Dimension Vista Akaline Phosphatase (ALPI) Flex reagent cartridge - In vitro diagnostic test for the quantitative measurement of alkaline phosphatase in human serum and plasma Catalog Number: K2115, SMN# 10642444
Low outlier resultsgenerated with Alkaline Phosphatase (ALPI) Flex reagent cartridge.The maximum difference observed with a patient sample was approximately -27% at a concentration of 129 U/L (2.15 ukat/L). The low outlier results are generated from the last five (5) tests of the well
Siemens issued Urgent Medical Device Recall (UMDC # VC-20-02.A.US via Fed'X) and Urgent Field Safety Notice (UFSN # VC-20-02.A.OUS) on August 25, 2020 to all Dimension Vista customers who received the Dimension Vista ALPI flex reagent lots 19247AB, 19282BB and 19330BD to notify them of the issue, health risk and actions to take. Customers were instructed to: Discontinue use of and discard the Dimension Vista ALPI lots 19247AB, 19282BB, 19330BD. Review their inventory of these lots to determine their laboratorys replacement needs. Complete and return the attached form to this letter to request your no-charge replacement product(s). Complete and return the Field Correction Effectiveness Check Form attached to the letter within 30 days. If they received any complaints of illness or adverse events associated with the product listed in the letter, to immediately contact their local Siemens Remote Services Center or their local Siemens Technical Support Representative. Expanded Recall 3/12/21: Urgent Medical Device Correction (UMDC # VC-20-02.B.US) and Urgent Field Safety Notice (UFSN # VC-20-02.B.OUS) issued on March 12, 2021 to all Dimension Vista customers who received the Dimension Vista ALPI flex reagent lots 20337AB, 20311BA, 20275AA, 20218AA, 20203BA, 20170AA, 20125BA and 20078AB to notify them of the issue, and actions to be taken.
Worldwide distribution - US Nationwide and the countries of Canada, Australia, Austria, Belgium, Bahamas, Czech Republic, Denmark, France, Germany, Italy, Japan, Korea, Netherlands, Norway, New Zealand, Portugal, Slovakia, Slovenia, Spain, Switzerland.
US:5970 units ; OUS: 1446 units