Baxter Flo-Gard 6301 Dual Channel Volumetric Infusion Pump; Baxter Healthcare Corporation, Deerfield, IL 60015; Product Code # 2M8064. The Flo-Gard Volumetric Infusion Pumps are an electromechanical pump used for intravenous infusion of fluids at user-selected rates.
Recall
- Recall Number
- Z-0134-2012
- Event Number
- 60168
- Firm
- Baxter Healthcare Corp. Rt.
- FEI Number
- 1416980
- Product Code
- FRN
- Status
- Terminated
- Root Cause
- Employee error
- Initiated
- August 26, 2011
- Posted
- October 28, 2011
- Terminated
- October 31, 2011
- Address
- 120 & Wilson Rd, Round Lake, IL, 60073
Description
Baxter Flo-Gard 6301 Dual Channel Volumetric Infusion Pump; Baxter Healthcare Corporation, Deerfield, IL 60015; Product Code # 2M8064. The Flo-Gard Volumetric Infusion Pumps are an electromechanical pump used for intravenous infusion of fluids at user-selected rates.
The Flo-Gard infusion Pump was released to the customer with OOS values for the Air Sensor Calibrations with the Air in Tubing test.
A Baxter Healthcare technician initially notified the sole customer on August 26, 2011 of the defective product and requested that the product be quarantined. On September 14, 2011 a technician visited the site to correct the problem observed on the Flo-Gard Infusion Pump. All required tests were performed and the results were found acceptable.
Distributed only in Puerto Rico.
1 unit