Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0740-140-2, NG-NV-7-40, NG-0640-140-2, NG-NV-6-40
Recall
- Recall Number
- Z-0131-2026
- Event Number
- 97522
- Firm
- Contego Medical LLC
- FEI Number
- 3011471056
- Product Code
- NIM
- Status
- Open, Classified
- Root Cause
- Device Design
- Initiated
- September 4, 2025
- Posted
- October 10, 2025
- Address
- 3801 Lake Boone Trl, Ste 100, Raleigh, NC, 27607-2994
Description
Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0740-140-2, NG-NV-7-40, NG-0640-140-2, NG-NV-6-40
Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter that may not open or may not fully open during sheath retraction and operation of the filter knob. The stent is used with a primary distal embolic protection device.
On 9/4/2025, communication and training to the field representatives started with in person letter distribution following immediately after. Customers were asked to do the following: Action Required 1) Inventory Check: Medtronic and Contego sales representatives will visit your facility to conduct an inventory count of devices from the affected lots. 2) Quarantine Affected Devices: Any identified devices will be removed from use and quarantined by the Medtronic or Contego team. 3) Return: Medtronic and Contego will remove affected devices. If replacements are desired, they can be provided during the same visit. 4) Confirmation: Sign and return the enclosed acknowledgment form to confirm receipt of this advisory and the completion of actions performed by Medtronic or Contego. 5) Please email the completed form to Contego Medical [email protected] as soon as possible and retain a copy of the form to provide to the representative at the time of their visit. If you have questions, contact the firm's Recall Support Team at [email protected] or call 919-459-7250 Extension 4.
US Nationwide distribution in the states of FL, CO, CA, IL, NC, KY, TX, PA, DE, NJ, IN, NY, OH, LA, OK, SC, CT, WI, MA, KS, UT, MD, AL.