This device is not approved in the United States. Medtronic Virtuoso VR, D164VWC. Medtronic, Inc., Minneapolis, MN 55432 USA.

Recall: Z-0131-2010 · Initiated September 9, 2009

Recall

Recall Number
Z-0131-2010
Event Number
53294
Firm
Medtronic Inc. Cardiac Rhythm Disease Managment
FEI Number
2182208
Product Code
LWS
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
September 9, 2009
Posted
November 9, 2009
Terminated
May 29, 2012
Address
8200 Coral Sea St. N.E., Saint Paul, MN, 55112

Description

This device is not approved in the United States. Medtronic Virtuoso VR, D164VWC. Medtronic, Inc., Minneapolis, MN 55432 USA.

Reason

A subset of active implanted Concerto CRT-D and Virtuoso ICD devices worldwide (estimated at 6,300) may not meet expected device longevity due to gradually increasing current drain caused by low voltage capacitor degradation. This issue may present in the affected devices as reaching the Recommended Replacement Time (RRT) earlier than projected. This issue does not compromise device functionalit

Action

An "Important Medical Device Information" letter, dated September 2009 was hand delivered by Medtronic Representatives beginning 09/09/09. The letter describes the issue, identifies affected devices and provides recommendations for patients with devices in the affected subset.. A Patient letter was mailed to patients beginning 09/28/09. For additional information: Doctors are advised to contact your local Medtronic Representative or Medtronic Technical Services at 800-723-4636; Device tracking managers are advised to contact your Medtronic representative, or the Patient Registration Department at 1-800-328-2518 x41832; and Patients are advised to contact Medtronic Patient Services number at 800-551-5544 ext. 41835.

Distribution

Worldwide distribution (except Idaho and Wyoming) including countries of Australia, Austria, Belgium, Canada, Czech Republic, Denmark, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Kuwait, Netherlands, Norway, Portugal, Saudi Arabi, Spain, Sweden, Switzerland, Turkey and United Kingdom.

Quantity

405