American Medical Systems, Inc., (AMS) GreenLight HPS Fiber Optic (ADDStat) Part Number 0010-2090B. Fiber optic laser for treatment of BPH (Benign Prostatic Hyperplasia). AMS Innovative Center, San Jose, CA 95134
Recall: Z-0131-2008
·
Initiated June 18, 2007
Recall
- Recall Number
- Z-0131-2008
- Event Number
- 38377
- Firm
- AMS Innovative Center San Jose
- FEI Number
- 2937094
- Product Code
- GEX
- Status
- Terminated
- Root Cause
- Process design
- Initiated
- June 18, 2007
- Posted
- February 26, 2008
- Terminated
- March 23, 2010
- Address
- 3070 Orchard Dr, San Jose, CA, 95134-2011
Description
American Medical Systems, Inc., (AMS) GreenLight HPS Fiber Optic (ADDStat) Part Number 0010-2090B. Fiber optic laser for treatment of BPH (Benign Prostatic Hyperplasia). AMS Innovative Center, San Jose, CA 95134
Reason
Control knob break: The knob used to finely control the direction of the laser energy may break.
Action
American Medical Systems, Inc. (AMS) issued Important: Medical Device Recall AMS Greenlight ADDStat Fibers notification letters to customers who have purchased the product on June 28, 2007. The customer letter requests recipients that the product use be discontinued, and the fibers be returned. The firm will send replacements.
Distribution
Worldwide Distribution - USA and Singapore.
Quantity
611 units