GreenLight HPS System, Surgical Laser System and Accessories. Part Numbers 0010-0070 and 0010-0075 (international units), AMS Innovative Center, San Jose, CA 95134.
Recall
- Recall Number
- Z-0130-2008
- Event Number
- 38336
- Firm
- AMS Innovative Center San Jose
- FEI Number
- 2937094
- Product Code
- GEX
- Status
- Terminated
- Root Cause
- Process design
- Initiated
- April 17, 2007
- Posted
- March 20, 2008
- Terminated
- March 23, 2010
- Address
- 3070 Orchard Dr, San Jose, CA, 95134-2011
Description
GreenLight HPS System, Surgical Laser System and Accessories. Part Numbers 0010-0070 and 0010-0075 (international units), AMS Innovative Center, San Jose, CA 95134.
Component missing: Some products were issued without thermal protection switches.
AMS issued Important: Medical Device Correction letters, dated April 17, 2007, to its US and international consignees by mail, instructing them to return an acknowledgement form. The letter informs customers that some HPS systems may lack a motor overload switch, that the firm believes that there is no immediate safety issue, and installation of the MOL switch will be conducted during normal preventive maintenance. AMS expects to complete installation of the switches by end of August 2007.
Worldwide Distribution - USA, Italy, France, Germany, Spain, UK, Denmark, Sweden, Australia, India, Japan, Greece, Switzerland, Saudi Arabia, Thailand, Turkey, Singapore, Taiwan/Hong Kong (China), and Austria.
169 units