FDA Recall
Terminated
Dimension Vista HbA1c
Recall: Z-0127-2015
·
Initiated August 14, 2014
Recall
- Recall Number
- Z-0127-2015
- Event Number
- 69334
- Firm
- Siemens Healthcare Diagnostics, Inc.
- FEI Number
- 2517506
- Product Code
- LCP
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- August 14, 2014
- Posted
- October 24, 2014
- Terminated
- August 10, 2015
- Address
- 500 Gbc Dr PO BOX 6101, Ms 514, Newark, DE, 19702-2466
Description
Dimension Vista HbA1c
Reason
In March, Siemens Healthcare Diagnostics issued an Urgent Medical Device Correction for decreased recovery of patient samples with the Dimension Vista® CSA & CSAE Flex® reagent cartridges when whole blood samples are run from Small Sample Containers (SSC). Siemens has confirmed that when whole blood samples are run from SSC, HbA1c values may exhibit a percent bias ranging from - 8.2% t
Action
An Urgent Medical Device Correction, dated August 2014, was provided to all Dimension Vista HbA1c customers who received the affected lots to notify them of the issue, the potential risk to health, and actions to be taken by customers.
Distribution
Worldwide Distribution: US (nationwide) and country of: Canada.
Quantity
20043