FDA Recall Terminated

Dimension Vista HbA1c

Recall: Z-0127-2015 · Initiated August 14, 2014

Recall

Recall Number
Z-0127-2015
Event Number
69334
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2517506
Product Code
LCP
Status
Terminated
Root Cause
Device Design
Initiated
August 14, 2014
Posted
October 24, 2014
Terminated
August 10, 2015
Address
500 Gbc Dr PO BOX 6101, Ms 514, Newark, DE, 19702-2466

Description

Dimension Vista HbA1c

Reason

In March, Siemens Healthcare Diagnostics issued an Urgent Medical Device Correction for decreased recovery of patient samples with the Dimension Vista® CSA & CSAE Flex® reagent cartridges when whole blood samples are run from Small Sample Containers (SSC). Siemens has confirmed that when whole blood samples are run from SSC, HbA1c values may exhibit a percent bias ranging from - 8.2% t

Action

An Urgent Medical Device Correction, dated August 2014, was provided to all Dimension Vista HbA1c customers who received the affected lots to notify them of the issue, the potential risk to health, and actions to be taken by customers.

Distribution

Worldwide Distribution: US (nationwide) and country of: Canada.

Quantity

20043