Carestream DRX-Revolution Mobile X-ray System. Produces radiographic images.

Recall: Z-0126-2014 · Initiated June 6, 2013

Recall

Recall Number
Z-0126-2014
Event Number
65469
Firm
Carestream Health, Inc.
FEI Number
1315356
Product Code
IZL
Status
Terminated
Root Cause
Device Design
Initiated
June 6, 2013
Posted
October 31, 2013
Terminated
May 20, 2016
Address
150 Verona Street, Rochester, NY, 14608-1733

Description

Carestream DRX-Revolution Mobile X-ray System. Produces radiographic images.

Reason

Carestream Health Inc. is recalling the power plug on their DRX-Revolution Mobile X-Ray System due to the item generating sparks and becoming warm to touch.

Action

On June 6, 2013 Carestream Health Inc. initiated a recall for the power plug on their DRX-Revolution Mobile X-Ray System due to the item generating sparks and becoming warm to touch. Carestream Health Inc. will replace the power plug in all potentially impacted DRX-Revolution Mobile X-Ray Systems. Your Carestream service representative will contact you to answer any questions and schedule a visit to replace the plug of your Mobile X-Ray System as soon as possible. If you have any questions or concerns, or if you experience any issue with your plug prior to our visit, please contact the Carestream Customer Care Center at 1-800-328-2910, 8AM through 8pm Eastern Standard Time.

Distribution

Nationwide Distribution including the states of NY, TX, PA, OH, WI, CA, IL, CO, FL, IN, OK, NJ, GA, LA, AL, KS, AZ, MA, SC, MI, NC, IA, UT, VA, TN, and MN.

Quantity

180 units