FDA Recall Terminated

Allura Xper FD10/10 Biplane X-ray system. Intended for use in cardiovascular and vascular x-ray imaging applications, including diagnostic, interventional procedures (such as PTCA and stent placement and atherectomies), pacemaker implantations and electrophysiology. It is compatible with specified magnetic navigation systems.

Recall: Z-0125-2011 · Initiated August 21, 2008

Recall

Recall Number
Z-0125-2011
Event Number
49488
Firm
Philips Medical Systems North America Co. Phillips
FEI Number
1000524572
Product Code
IZI
Status
Terminated
Root Cause
Software design
Initiated
August 21, 2008
Posted
October 22, 2010
Terminated
November 19, 2010
Address
22100 Bothell Everett Hwy, Bothell, WA, 98021-8431

Description

Allura Xper FD10/10 Biplane X-ray system. Intended for use in cardiovascular and vascular x-ray imaging applications, including diagnostic, interventional procedures (such as PTCA and stent placement and atherectomies), pacemaker implantations and electrophysiology. It is compatible with specified magnetic navigation systems.

Reason

Potential for table and stand not to move due to a software error.

Action

Philips Medical Systems North America Co. sent an URGENT - Device Correction, dated September 15, 2010, to their consignees. The letter identified the product, the problem, and the action to be taken. The consignees were informed that there was a potential geometry software lock up during a Digital Rotational Angiography (DRA) run, the stand and table movement may no longer be possible. The consignees were advised that as geometry software errors happen intermittently, no actions can be performed by the consignees to minimize the risk except for not performing DRA procedures until the firm upgrades the software. For question consignees were to contact the Philips Call Center at 800-722-9377 #5, #4, #1 and reference FCO 72200111 or FCO 72200125.

Distribution

Devices were distributed to hospitals and medical facilities through out the U.S.

Quantity

26 units