Philips PageWriter Touch Cardiograph, Model: 860284, Philips Medical Systems, Andover, MA, 01810
Recall
- Recall Number
- Z-0124-2008
- Event Number
- 44824
- Firm
- Philips Medical Systems
- FEI Number
- 1218950
- Product Code
- DPS
- Status
- Terminated
- Root Cause
- Software change control
- Initiated
- September 12, 2007
- Posted
- January 30, 2008
- Terminated
- August 16, 2016
- Address
- 3000 Minuteman Road, Andover, MA, 01810
Description
Philips PageWriter Touch Cardiograph, Model: 860284, Philips Medical Systems, Andover, MA, 01810
Incorrect display: When using the 16 lead Patient Interface Module (PIM), lead tracings of V3R and V4R are reversed in the extended lead modes of Pediatric and Balanced.
Philips Medical Systems notified affected customers by Customer Letter and Urgent - Device Correction Notice dated 9/12/07. The correction will be a software upgrade. In the US, the Philips Response Center will contact US customers and arrange their upgrades, either via a software download, or by sending a Field Service Engineer to the customer site to perform the upgrade. Until new software is installed, customers are being advised to use the 12 lead module, and to disable or disconnect the 16 lead connection.
Worldwide Distribution - USA, Canada, Australia, Hungary, New Zealand, Nigeria, Norway, Qatar, Taiwan, and UAE.
266 units