UniCel DxC 600/800 with Software v5.0.11, Part Number B12218. The UniCel DxC Synchron Clinical Systems are fully automated, computercontrolled clinical chemistry analyzers designed for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries. Analysis can be performed on serum, plasma, urine, or cerebrospinal fluid (CSF) and whole blood (sample type is chemistry dependent).
Recall
- Recall Number
- Z-0123-2013
- Event Number
- 63143
- Firm
- Beckman Coulter Inc.
- FEI Number
- 2050012
- Product Code
- JJE
- Status
- Terminated
- Root Cause
- Software change control
- Initiated
- September 3, 2012
- Posted
- October 24, 2012
- Terminated
- February 14, 2014
- Address
- 250 S Kraemer Blvd, Brea, CA, 92821-6232
Description
UniCel DxC 600/800 with Software v5.0.11, Part Number B12218. The UniCel DxC Synchron Clinical Systems are fully automated, computercontrolled clinical chemistry analyzers designed for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries. Analysis can be performed on serum, plasma, urine, or cerebrospinal fluid (CSF) and whole blood (sample type is chemistry dependent).
Beckman Coulter has confirmed that the Shuttle to Barcode Reader and Cap Piercer alignment procedure cannot be performed on certain DxC configurations with v5.0.11 software installed. If the Shuttle to Barcode Reader becomes misaligned, the Operator will not be able to perform this alignment procedure on the DxC systems listed above. Software v5.0 .11 does not allow access to the Shuttle to Bar
Beckman Coulter sent an Urgent Field Safety Notice dated August 30, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed that their Beckman Coulter Service Representative would contac them to reinstall the software. Customers were asked to complete and return the enclosed Response Form. For questions customers were instructed to call their Customers Support Center on the Beckman Coulter website at www.BeckmanCoulter.com/CustomerSupport/SupportRequest. For questions regarding this recall call 714-993-5321.
Worldwide Distribution - USA including TX and OH. Interntionally to China and Turkey
7 units total (both US and International)