FDA Recall Open, Classified

PERIFIX FX Continuous Epidural Anesthesia Set, PERIFIX Tuohy Epidural Needle - 17 Ga. x 3-1/2 in. (8.9 cm) - Winged, PERIFIX FX Springwound Epidural Catheter - 19 Ga. Closed tip. Product Code: CE17TFC. Product Description: CE17TFC CONTINUOUS EPID SET W/17G TUOHY.

Recall: Z-0121-2026 · Initiated September 18, 2025

Recall

Recall Number
Z-0121-2026
Event Number
97550
Firm
B BRAUN MEDICAL INC
FEI Number
3013764800
Product Code
OGE
Status
Open, Classified
Root Cause
Process control
Initiated
September 18, 2025
Posted
October 9, 2025
Address
861 Marcon Blvd, Allentown, PA, 18109-9577

Description

PERIFIX FX Continuous Epidural Anesthesia Set, PERIFIX Tuohy Epidural Needle - 17 Ga. x 3-1/2 in. (8.9 cm) - Winged, PERIFIX FX Springwound Epidural Catheter - 19 Ga. Closed tip. Product Code: CE17TFC. Product Description: CE17TFC CONTINUOUS EPID SET W/17G TUOHY.

Reason

Potential for the lid of the catheter connector to be in the incorrect position.

Action

B Braun notified consignees on about 09/18/2025 via USPS certified mailed URGENT MEDICAL DEVICE CORRECTION NOTIFICATION. Consignees were instructed to, where possible, have a second kit or additional catheter connector available at the point of use and upon opening the kit, verify the correct catheter connector is present. If the affected component is detected, discard and do not use. Consignees were also instructed to review the Medical Device Correction Notification and ensure all users in your organization and other concerned persons are informed, if affected units have been further distributed, notify customers, post the notification where affected units are stored, and complete and return the "Urgent Medical Device Correction Acknowledgement Form." Additionally, if consignees have identified affected units in inventory, they may order one case (50 pieces) of affected sterilized single use catheter connectors to have on hand.

Distribution

Nationwide distribution to AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV.

Quantity

0 units