P2131 Automated Field Steam Sterilizer NSN: 6530-01-641-4641
Recall
- Recall Number
- Z-0121-2022
- Event Number
- 88628
- Firm
- Fort Defiance Industries, LLC
- FEI Number
- 3011421558
- Product Code
- FLE
- Status
- Open, Classified
- Root Cause
- Device Design
- Initiated
- August 19, 2021
- Address
- 2411 Maremont Pkwy, Loudon, TN, 37774-1066
Description
P2131 Automated Field Steam Sterilizer NSN: 6530-01-641-4641
The rear heater plate and cover assembly that contains the electrical wiring for the heaters does not have sufficient bonding to ground.
A Recall notification letter titled, "MEDICAL DEVICE ADVISORY NOTICE" was sent to consignees on 08/19/2021 via a email. The notice provides instructions regarding continued use of the device until FDI provides a protective cover for the rear heater plate and cover assembly. It mentions, "THE OPERATOR SHOULD NOT TOUCH THE REAR HEATER PLATE AND COVER ASSEMBLY OF THE DEVICE WHILE THE DEVICE IS IN USE OR POWERED ON. IF ACCESS TO THE REAR OF THE DEVICE IS REQUIRED, PLEASE TURN THE POWER OF THE UNIT OFF USING THE FRONT PANEL SWITCH OR DISCONNECT INCOMING POWER TO THE DEVICE. DO NOT ATTEMPT TO MODIFY THE DEVICE TO CHANGE THE ELECTRICAL GROUND CONDITION OF THE HEATER PLATE AND COVER ASSEMBLY AS THIS WILL IMPACT DEVICE PERFORMANCE."
Distributed nationwide to PA, MD, TX, NJ, OH, TN, UT, VA and internationally to Canada.
380 devices