FDA Recall Terminated

Xario XG Diagnostic Ultrasound System SSA-680A Used for routine diagnostic imaging

Recall: Z-0120-2014 · Initiated August 5, 2013

Recall

Recall Number
Z-0120-2014
Event Number
66059
Firm
Toshiba American Medical Systems Inc
FEI Number
2020563
Product Code
IYN
Status
Terminated
Root Cause
Device Design
Initiated
August 5, 2013
Posted
October 30, 2013
Terminated
June 28, 2016
Address
2441 Michelle Dr P.O. Box 2068, Tustin, CA, 92780-7047

Description

Xario XG Diagnostic Ultrasound System SSA-680A Used for routine diagnostic imaging

Reason

The firm initiated this recall of certain Toshiba Diagnostic Ultrasound Systems because it has been found that under certain conditions, the temperature of the transducer head may increase in 2D+PW mode or 2D+CDI (POWER, ADF) +PW mode. When the system is used under both conditions listed below, the temperature of the transducer head connected to the system may increase in temperature and possibly

Action

Toshiba sent an Urgent Medical Device Correction letter dated August 5, 2013. along with a Reply Form via USPS return receipt mail. The letter identified the product, the problem, and the action to be taken by the customer. Until the software is corrected customers were instructed to change the preset settings as described in the letter. Customers were asked to complete and return the attached form and fax it to the toll free number. The form also can be sent via email to [email protected] Customers with questions were instructed to call 800-421-1968 or contact their local Toshiba Representative at 800-521-1968. For questions regarding this recall call 714-730-5000.

Distribution

Nationwide Distribution including and US Territory Puerto Rico

Quantity

477 (1352 total)