FDA Recall
Terminated
Michigan Instruments Thumper Mechanical CPR Device; Model 1007.
Recall: Z-0119-2007
·
Initiated August 4, 2006
Recall
- Recall Number
- Z-0119-2007
- Event Number
- 35934
- Firm
- Michigan Instruments, Inc.
- FEI Number
- 1821850
- Product Code
- DRM
- Status
- Terminated
- Root Cause
- Other
- Initiated
- August 4, 2006
- Posted
- November 3, 2006
- Terminated
- December 6, 2006
- Address
- 4717 Talon Court S.E., Grand Rapids, MI, 49512-5408
Description
Michigan Instruments Thumper Mechanical CPR Device; Model 1007.
Reason
Failure to initiate compressions when first turned on, if improperly shut down-Operator Manual updated to include proper shut down procedures
Action
Consignees were sent a letter dated 8/4/06 with additional instructions for use. A second letter was sent on 9/1/06 to correct the new instructions provided in the first letter for page 19 of the instructions.
Distribution
Worldwide- USA and countries of Brazil, Canada, China, Germany, Greece, Japan, Korea, Macau, Netherlands, Pakistan, Spain, Taiwan and United Kingdom.
Quantity
1100