FDA Recall Terminated

3M Bair Hugger(TM) Normothermia System, Temperature Monitoring System Sensors Model 360 (Part Number 36000)

Recall: Z-0114-2018 · Initiated October 17, 2017

Recall

Recall Number
Z-0114-2018
Event Number
78327
Firm
3M Company Health Care Business 3m Center
FEI Number
2110898
Product Code
FLL
Status
Terminated
Root Cause
Software design
Initiated
October 17, 2017
Terminated
November 3, 2020
Address
2510 Conway Ave, B # 275-5-W-6, Saint Paul, MN, 55144-0001

Description

3M Bair Hugger(TM) Normothermia System, Temperature Monitoring System Sensors Model 360 (Part Number 36000)

Reason

During a recent investigation, 3M confirmed that a programming translation error could occur in a small amount of the sensors that could lead to a temperature readout that is lower than the patients actual temperature.

Action

The firm initiated their recall on 10/17/2017 via mail. International subsidiaries were notified via email, also on October 17, 2017.Distributors were asked to notify their customers and all direct consignees were directed to dispose of the product on site.

Distribution

CA, MI, IL, MO, MN, GA, and Canada, Switzerland

Quantity

153 cases (3825 sensors)