GE Healthcare Voluson E8 ultrasound scanner with Software version 6.2.0 and 6.2.1 and DICOM stations with these software versions; GE Healthcare 9900 West Innovation Drive, Wauwatosa, WI 53226, USA.
Recall: Z-0113-2008
·
Initiated July 6, 2007
Recall
- Recall Number
- Z-0113-2008
- Event Number
- 39400
- Firm
- Ge Healthcare
- FEI Number
- 3005860720
- Product Code
- IYO
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- July 6, 2007
- Posted
- November 7, 2007
- Terminated
- October 21, 2008
- Address
- 9900 W Innovation Dr, Wauwatosa, WI, 53226-4856
Description
GE Healthcare Voluson E8 ultrasound scanner with Software version 6.2.0 and 6.2.1 and DICOM stations with these software versions; GE Healthcare 9900 West Innovation Drive, Wauwatosa, WI 53226, USA.
Reason
When a spectrum of a Pulsed Wave Doppler shows an angle correction unequal to 0 is reloaded from the archive or the image is sent to a DICOM station, this angle correction is mistakenly disregarded resulting in an underestimation of the flow velocities.
Action
GE Healthcare sent customer letters on July 15, 2007 to affected customers. The letter included a short term mitigation recommendation and a long-term correction. The long-term correction will be software upgrade provided free of charge to all affected customers.
Distribution
Nationwide.
Quantity
521 Units