FDA Recall Terminated

GE Healthcare Voluson E8 ultrasound scanner with Software version 6.2.0 and 6.2.1 and DICOM stations with these software versions; GE Healthcare 9900 West Innovation Drive, Wauwatosa, WI 53226, USA.

Recall: Z-0113-2008 · Initiated July 6, 2007

Recall

Recall Number
Z-0113-2008
Event Number
39400
Firm
Ge Healthcare
FEI Number
3005860720
Product Code
IYO
Status
Terminated
Root Cause
Software design
Initiated
July 6, 2007
Posted
November 7, 2007
Terminated
October 21, 2008
Address
9900 W Innovation Dr, Wauwatosa, WI, 53226-4856

Description

GE Healthcare Voluson E8 ultrasound scanner with Software version 6.2.0 and 6.2.1 and DICOM stations with these software versions; GE Healthcare 9900 West Innovation Drive, Wauwatosa, WI 53226, USA.

Reason

When a spectrum of a Pulsed Wave Doppler shows an angle correction unequal to 0 is reloaded from the archive or the image is sent to a DICOM station, this angle correction is mistakenly disregarded resulting in an underestimation of the flow velocities.

Action

GE Healthcare sent customer letters on July 15, 2007 to affected customers. The letter included a short term mitigation recommendation and a long-term correction. The long-term correction will be software upgrade provided free of charge to all affected customers.

Distribution

Nationwide.

Quantity

521 Units