FDA Recall Terminated

MEDLINE PEDI ELBOW CLOSED SUCTION CATHETER, 8Fr, REF DYNCPE8

Recall: Z-0112-2018 · Initiated October 10, 2017

Recall

Recall Number
Z-0112-2018
Event Number
78319
Firm
Medline Industries Inc
FEI Number
1417592
Product Code
BSY
Status
Terminated
Root Cause
Employee error
Initiated
October 10, 2017
Posted
November 18, 2017
Terminated
August 4, 2020
Address
Three Lakes Drive, Northfield, IL, 60093

Description

MEDLINE PEDI ELBOW CLOSED SUCTION CATHETER, 8Fr, REF DYNCPE8

Reason

Products labeled as sterile were distributed prior to sterilization

Action

The recall was initiated by mail on 10/10/2017. The letter contained the following directive: "REQUIRED ACTION: 1. Immediately check your stock for item numbers DYNCPE8 (Lot 2861707008) & DYNCSDS12 (Lot 28617070015) place them on hold and complete the attached response form. Quarantine all affected product / lots. 2. Please return the completed enclosed response form listing the quantity of affected product on hand. Even if you do not have any affected product please complete and return the form, as indicated on the response form. When we receive your completed response form you will be sent return labels (if applicable). Your account will receive credit when the returned product is received. 3. If you have transferred possession of this product to another individual, to another department or another location in your organization you are required by FDA regulations to notify them of this recall communication. Have the customers return any affected product to you for collection and return to Medline Industries, Inc. If you have any questions, please contact 866-359-1704."

Distribution

FL, IN, TN, GA, NJ, UT, MD, Nicaragua, Honduras

Quantity

190 units