Siemens Ultrasound, Acuson/Sonovista X300 ultrasound systems, software revisions 2.0.1 to 2.0.05, 3.0.01 and 3.0.02, manufactured by Siemens Ultrasound

Recall: Z-0111-2009 · Initiated August 19, 2008

Recall

Recall Number
Z-0111-2009
Event Number
49412
Firm
Siemens Medical Solutions USA, Inc.
FEI Number
2936884
Product Code
IYN
Status
Terminated
Root Cause
Employee error
Initiated
August 19, 2008
Posted
October 10, 2008
Terminated
December 20, 2010
Address
1230 Shorebire Way P.O. Box 7393, Mountain View, CA, 94043

Description

Siemens Ultrasound, Acuson/Sonovista X300 ultrasound systems, software revisions 2.0.1 to 2.0.05, 3.0.01 and 3.0.02, manufactured by Siemens Ultrasound

Reason

Thermal Index cranial (TIC) is not displayed for the Neo-Head exam type with the C8-5 transducer. This is a required display to support the alara principle when imaging cranial structure.

Action

Field correction was initiated on 8/19/2008. Under Siemens update program US020/08/S & US021/08/S Siemens service personnel make a site visit, deliver a customer safety advisory notice to the customer and correct the problem by installing the revised Software versions. After the planned field correction is performed at the customer site the Service personnel log that information through Siemens internal web or reporting tools into the Siemens field update database.

Distribution

Worldwide Distribution --- including USA and countries of France, Colombia, Poland, Taiwan, Saudi Arabia, Austria, Malaysia, India, Czech Republic, Croatia, Argentina, Venezuela, Italy, Philippines, Mauritius, Spain, Bolivia, Germany, Ukrain, Russia, United Kingdom, Brazil, South Africa and Canada.

Quantity

72 units