FDA Recall Open, Classified

TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, D, (UniD Curve D), REF A-TFSE-D; cardiac catheter

Recall: Z-0109-2026 · Initiated September 10, 2025

Recall

Recall Number
Z-0109-2026
Event Number
97595
Firm
Abbott
FEI Number
3005334138
Product Code
OAE
Status
Open, Classified
Root Cause
Package design/selection
Initiated
September 10, 2025
Posted
October 10, 2025
Address
5050 Nathan Ln N, Plymouth, MN, 55442-3209

Description

TactiFlex Sensor Enabled Ablation Catheter, 8F, 115 cm, D, (UniD Curve D), REF A-TFSE-D; cardiac catheter

Reason

Abbott has become aware of a limited number of tip detachment events during catheter removal from the packaging.

Action

Abbott issued a Medical Device Advisory to its consignees on 09/10/2025 by letter delivered via Federal Express. The notice explained the issue, impact and associated risk to the patient and requested the following: "Action Requested/Recommendations/Next Steps . Continue following the product IFU by inspecting the packaging and catheter for damage and/or kinking prior to use. . Review the guidance in Appendix A information on proper removal technique of the TactiFlex" Ablation Catheter, Sensor Enabled" from the packaging tray, ensuring to inspect the electrodes and catheter tip carefully for integrity and overall condition prior to use. . Abbott has made changes to the packaging tray design to help mitigate tip damage during catheter removal from the packaging. The deployment of the new packaging has begun. Please complete the provided Acknowledgement Form with the assistance of your Abbott representative. This form ensures you are aware of the advisory notice and have been given an opportunity to ask questions. Please forward this notice to anyone in your organization who may need to be notified and maintain this notice for your records to ensure effectiveness of this communication. Should you have questions about this issue, please contact your local Abbott representative or Abbott Support at 1-855-478-5833 (U.S), 8:30 a.m. - 5:30 p.m. Central Standard Time, Monday thru Friday."

Distribution

Worldwide

Quantity

407 units