FDA Recall Terminated

SkyVision SDS System - Video Integration Model: GC-SDS-BSCB-200 - Product Usage: is a high quality, self-contained video switching and control unit for use in the operating room.

Recall: Z-0109-2021 · Initiated March 10, 2020

Recall

Recall Number
Z-0109-2021
Event Number
86439
Firm
Skytron, Div. The KMW Group, Inc
FEI Number
1825014
Product Code
LMD
Status
Terminated
Root Cause
Software Manufacturing/Software Deployment
Initiated
March 10, 2020
Terminated
June 1, 2021
Address
5085 Corporate Exchange Blvd SE, Grand Rapids, MI, 49512-5515

Description

SkyVision SDS System - Video Integration Model: GC-SDS-BSCB-200 - Product Usage: is a high quality, self-contained video switching and control unit for use in the operating room.

Reason

Intermittent failure of a signal being sent to a monitor from the touch panel display causing the displaying monitor to go Blank and the touch panel to display a Blank image when reviewing the target display. If this issue were to happen during a case, cause increase risk to patient due to hospital losing the visual image.

Action

Skytron Authorized Representatives contacted the affected facilities beginning on 3/4/20 via phone or in person .Skytron Authorized Representatives visited sites to correct the issue wih the following message: Skytron, LLC is voluntarily recalling a limited number of SDS integration units. The SDS router may experience an intermittent issue related to display output functionality. This failure poses a potential safety risk to patients. This firmware update is being conducted to correct the failure.

Distribution

US Nationwide distribution including in the states of CA, FL, IA, MI, MN, NY, PA, TX, WA.

Quantity

36 units