FDA Recall Open, Classified

Evita V500 Ventilator with Installed CO2 Measurement Option-Intended for the ventilation of adult,pediatric and neonatal patients. Catalog Number: 8416400

Recall: Z-0108-2022 · Initiated September 10, 2021

Recall

Recall Number
Z-0108-2022
Event Number
88635
Firm
Draeger Medical, Inc.
FEI Number
2517967
Product Code
CBK
Status
Open, Classified
Root Cause
Software design
Initiated
September 10, 2021
Address
3135 Quarry Rd, Telford, PA, 18969-1042

Description

Evita V500 Ventilator with Installed CO2 Measurement Option-Intended for the ventilation of adult,pediatric and neonatal patients. Catalog Number: 8416400

Reason

Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds

Action

Drager issued Urgent Medical Device Recall letter on 9/10/21 via FedEx with tracking (confirmation of delivery). Letter states reason for recall, health risk and action to take: The release of a corrected software is expected to be available in Q4 of 2021. Once the software is available, your local Drger Service representative will contact you to arrange a date for the software update to be performed free of charge. According to our records, at least one of the mentioned affected devices has been shipped to your facility. Please ensure that all users of the above-mentioned products and other concerned persons within your organization are made aware of this Urgent Medical Device Recall Notice. If you have provided the products to third parties, please forward a copy of this information. Please keep this information at least until the corrective measure has been completed. Please also complete and return the attached Customer Acknowledgment Card to confirm to us that you have received this information. If you have any questions regarding the operation of your Evita V500 or Babylog VN500, please contact Drger Service Technical Support between the hours of 8:00 AM 8:00 PM EST at 1-800-437-2437 (press 2 at the prompt, then 2, then 1).

Distribution

US Nationwide

Quantity

6,797 units