FDA Recall Open, Classified

FlexLab X (FXX) System. Potassium Test System. in vitro diagnostic

Recall: Z-0107-2026 · Initiated September 3, 2025

Recall

Recall Number
Z-0107-2026
Event Number
97565
Firm
Inpeco S.A. Via San Gottardo 10 Lugano Switzerland
FEI Number
3010825766
Product Code
CEM
Status
Open, Classified
Root Cause
No Marketing Application
Initiated
September 3, 2025
Posted
October 9, 2025

Description

FlexLab X (FXX) System. Potassium Test System. in vitro diagnostic

Reason

The FlexLab X System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the FDA, so its use may cause an increase in erroneous results or a delay in patient results.

Action

Consignees received an Urgent Medical Device Recall notification via email dated 9/3/25. The notification instructs consignees to not rely on HIL functionality for clinical decision making or to flag potentially erroneous samples. HIL functionality must be discontinued and cannot be disabled by users. Inspeco must send an Inspeco Servicer to consignee locations to disable HIL functionality. Inspeco will contact consingnees to schedule this service. Consignees are to forward the provided notification to those whom it might concern. Consignees with questions can contact Regulatory Affairs via email at [email protected] or Inspeco Service Team via email at [email protected].

Distribution

US distribution to CA & NY.

Quantity

1 unit