FlexLab X (FXX) System. Potassium Test System. in vitro diagnostic
Recall
- Recall Number
- Z-0107-2026
- Event Number
- 97565
- Firm
- Inpeco S.A. Via San Gottardo 10 Lugano Switzerland
- FEI Number
- 3010825766
- Product Code
- CEM
- Status
- Open, Classified
- Root Cause
- No Marketing Application
- Initiated
- September 3, 2025
- Posted
- October 9, 2025
Description
FlexLab X (FXX) System. Potassium Test System. in vitro diagnostic
The FlexLab X System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the FDA, so its use may cause an increase in erroneous results or a delay in patient results.
Consignees received an Urgent Medical Device Recall notification via email dated 9/3/25. The notification instructs consignees to not rely on HIL functionality for clinical decision making or to flag potentially erroneous samples. HIL functionality must be discontinued and cannot be disabled by users. Inspeco must send an Inspeco Servicer to consignee locations to disable HIL functionality. Inspeco will contact consingnees to schedule this service. Consignees are to forward the provided notification to those whom it might concern. Consignees with questions can contact Regulatory Affairs via email at [email protected] or Inspeco Service Team via email at [email protected].
US distribution to CA & NY.
1 unit