Mindray V21, Size: 515 mm X 335 mm X 685 mm, N.W.: 8 kg, G.W.: 11 kg, Qty:1 The V-Series Monitor is intended for intra hospital use under the direct supervision of a licensed healthcare practitioner. The indications for use include the monitoring of ECG waveform derived from 3,5,6 and 12-lead measurements, Heart Rate, Pulse Oximetry (Sp)s), ST Segment Analysis, Arrhythmia Detection, Non Invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Cardia Output (CO), Respiratory Gases, Respiration Rate and Temperature.
Recall
- Recall Number
- Z-0107-2015
- Event Number
- 69260
- Firm
- Mindray DS USA, Inc. dba Mindray North America
- FEI Number
- 2221819
- Product Code
- MHX
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- July 31, 2014
- Posted
- October 21, 2014
- Terminated
- April 20, 2017
- Address
- 800 MacArthur Blvd, Mahwah, NJ, 07430-2001
Description
Mindray V21, Size: 515 mm X 335 mm X 685 mm, N.W.: 8 kg, G.W.: 11 kg, Qty:1 The V-Series Monitor is intended for intra hospital use under the direct supervision of a licensed healthcare practitioner. The indications for use include the monitoring of ECG waveform derived from 3,5,6 and 12-lead measurements, Heart Rate, Pulse Oximetry (Sp)s), ST Segment Analysis, Arrhythmia Detection, Non Invasive Blood Pressure (NIBP), Invasive Blood Pressure (IBP), Cardia Output (CO), Respiratory Gases, Respiration Rate and Temperature.
Mindray DS USA Inc. initiated a voluntary field corrective action for V-Series Patient Monitor because there is an issue with the V-Series Drug Calculator function.
Mindray DS USA, Inc. sent a Recall Correction Action Letter to their customers via certified mail with return receipt on July 31, 2014.
Worldwide Distribution: US (nationwide) and internationally to: Austalia, Canada, Columbia, Germany, India, Indonesia, Ireland, Israel, Kuwait, Malaysia, New Zealand, Pakistan, Russian Federation, Saudi Arabia, South Africa, Sri Lanka, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Venezula.
2,247 units