Siemens RAPIDLab 1245 system blood gas analyzer Product Usage: RAPIDLab 1200 Systems: The Rapidlab 1200 systems are intended for in vitro diagnostic use by healthcare professionals in the quantitative testing of human whole blood. The systems can determine the following parameters: RAPIDLab 1240 pH, pC02, p02 RAPIDLab 1245 pH, pC02, p02, tHb, nBili, F02Hb, FCOHb, FMetHb, FHHb RAPIDLab 1260 pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate RAPIDLab 1265 pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, nBili, F02Hb, FCOHb, FMetHb, FHHb
Recall
- Recall Number
- Z-0106-2014
- Event Number
- 66166
- Firm
- Siemens Healthcare Diagnostics Inc
- FEI Number
- 3002637618
- Product Code
- CHL
- Status
- Terminated
- Root Cause
- Software Manufacturing/Software Deployment
- Initiated
- September 17, 2012
- Posted
- October 29, 2013
- Terminated
- April 1, 2015
- Address
- 2 Edgewater Dr, Norwood, MA, 02062-4637
Description
Siemens RAPIDLab 1245 system blood gas analyzer Product Usage: RAPIDLab 1200 Systems: The Rapidlab 1200 systems are intended for in vitro diagnostic use by healthcare professionals in the quantitative testing of human whole blood. The systems can determine the following parameters: RAPIDLab 1240 pH, pC02, p02 RAPIDLab 1245 pH, pC02, p02, tHb, nBili, F02Hb, FCOHb, FMetHb, FHHb RAPIDLab 1260 pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate RAPIDLab 1265 pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, nBili, F02Hb, FCOHb, FMetHb, FHHb
RAPIDPoint 400 Series, RAPIDPoint 500, or RAPIDLab 1200 Series blood gas analyzer, users can inadvertently enter a wrong patient ID on the demographics screen, which ultimately could lead to a misidentification of the sample
Siemens Healthcare sent an Urgent Field Safety Notification letter dated September 2013 via Federal Express, to affected Siemens domestic customers. Effected Siemens customers outside the US was provided a copy of the Urgent Field Safety Notice via hard copy, e-mail, and/or fax as determined by each countries local regulations and procedures. Previously Siemens Healthcare had issued a Customer Notification on August 23, 2012. The letter identified the affected product, issue, information to mitigate the issue and actions to be taken. The firm will issue an upgrades software to users. For questions contact your local Customer Care Center - Technical Support Representative at 1-877-229-3711.
Worldwide Distribution - USA Nationwide and in the countries of Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia Herzeg., Botswana, Brazil, Brunei, Bulgaria, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Fren.Polynesia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Ivory Coast, Japan, Kazakhstan, Kenya, Kuwait, Latvia, Lithuania, Luxembourg, Macedonia, Malaysia, Mayotte, Netherlands, New Caledonia, New Zealand, Norway, P.R. China, Philippines, Poland, Portugal, Qatar, Rep. of Yemen, Republic Korea, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, U.A.E., United Kingdom, Vatikancity, Venezuela, Vietnam,
274 units