FDA Recall
Terminated
Multirall(TM) 200Liko Strap Gripper P/N 3136250 overhead lift for patient transfer situations.
Recall: Z-0105-2019
·
Initiated July 17, 2018
Recall
- Recall Number
- Z-0105-2019
- Event Number
- 80911
- Firm
- Liko AB Alvik
- FEI Number
- 6884
- Product Code
- FSA
- Status
- Terminated
- Root Cause
- Storage
- Initiated
- July 17, 2018
- Terminated
- April 19, 2019
- Address
- 7555 Lulea Sweden
Description
Multirall(TM) 200Liko Strap Gripper P/N 3136250 overhead lift for patient transfer situations.
Reason
The products manufactured on 5/5/17 may be assembled incorrectly. If the incorrectly assembled products is used for patient transport, this could allow the patient to fall, resulting in potential injury to the patient.
Action
The customer was contacted via telephone on 7/17/18 to notify them of the recall. An Urgent Field Safety Notice notification letter was sent to the customer on 7/19/18.
Distribution
The products were distributed to the following US states: OH.
Quantity
2