Sopha Medical Vision DST / DST-XL Digital vatiable angle Nuclear Imaging System

Recall: Z-0105-06 · Initiated September 27, 2005

Recall

Recall Number
Z-0105-06
Event Number
33832
Firm
General Electric Med Systems LLC
FEI Number
2126677
Product Code
KPS
Status
Terminated
Root Cause
Other
Initiated
September 27, 2005
Posted
October 28, 2005
Terminated
December 26, 2006
Address
3000 N Grandview Blvd, Waukesha, WI, 53188-1615

Description

Sopha Medical Vision DST / DST-XL Digital vatiable angle Nuclear Imaging System

Reason

Cracked welds on the main rotation shaft could potentially lead to the collapse of the gantry yokes that support the detectors if the system is continued to be used after the weld has completely fractured.

Action

GE Healthcare issued an Urgent Safety Notice, dated 07/22/05, to all affected sites. This letter describes the potential hazard, its waring signs and the additional corrective actions that will be taken by GE Healthcare.

Distribution

Within the US to include: AL, AZ, CA, DE, FL, GA, IA, IL, KS, MA, MD, MI, NC, NJ, NY, OH, PA, SC, TX, VA, WI and WV. OUS to include: Belgium, Brazil, Canada, China, Czech Republic, Egypt, France, Germany, Greece, Italy, Korea, Netherlands, Norway, Portugal, Russia, Saudi Arabia, Sinapore, Spain, Sweden and United Kingdom

Quantity

140 (80 US and 60 OUS)