Sopha Medical Vision DST / DST-XL Digital vatiable angle Nuclear Imaging System
Recall
- Recall Number
- Z-0105-06
- Event Number
- 33832
- Firm
- General Electric Med Systems LLC
- FEI Number
- 2126677
- Product Code
- KPS
- Status
- Terminated
- Root Cause
- Other
- Initiated
- September 27, 2005
- Posted
- October 28, 2005
- Terminated
- December 26, 2006
- Address
- 3000 N Grandview Blvd, Waukesha, WI, 53188-1615
Description
Sopha Medical Vision DST / DST-XL Digital vatiable angle Nuclear Imaging System
Cracked welds on the main rotation shaft could potentially lead to the collapse of the gantry yokes that support the detectors if the system is continued to be used after the weld has completely fractured.
GE Healthcare issued an Urgent Safety Notice, dated 07/22/05, to all affected sites. This letter describes the potential hazard, its waring signs and the additional corrective actions that will be taken by GE Healthcare.
Within the US to include: AL, AZ, CA, DE, FL, GA, IA, IL, KS, MA, MD, MI, NC, NJ, NY, OH, PA, SC, TX, VA, WI and WV. OUS to include: Belgium, Brazil, Canada, China, Czech Republic, Egypt, France, Germany, Greece, Italy, Korea, Netherlands, Norway, Portugal, Russia, Saudi Arabia, Sinapore, Spain, Sweden and United Kingdom
140 (80 US and 60 OUS)