FDA Recall Open, Classified

Multichem S Plus, Part Codes: a) 08P88-10, b) 08P88-11, c) 08P88-12

Recall: Z-0104-2024 · Initiated August 25, 2023

Recall

Recall Number
Z-0104-2024
Event Number
93017
Firm
TECHNO-PATH MANUFACTURING LTD. Fort Henry Business Park Ballina Tipperary Ireland
FEI Number
3008156187
Product Code
JJY
Status
Open, Classified
Root Cause
Unknown/Undetermined by firm
Initiated
August 25, 2023
Posted
October 12, 2023

Description

Multichem S Plus, Part Codes: a) 08P88-10, b) 08P88-11, c) 08P88-12

Reason

Technopath Manufacturing Ltd. has identified an issue with the glass vial commodity used to fill the Multichem S Plus Level 1, Multichem S Plus Level 2, Multichem S Plus Level 3, Multichem IA Plus and Multichem P Controls. Some customers have experienced broken vials when receiving or thawing Technopath Multichem IA Plus and Multichem S Plus Controls. Breakages have been observed upon delivery, while thawing or while opening the vial. The breakages can result in the bottom of the vial breaking off, breakage in the neck thread area or cracking on the sides of the vial. This may expose the user to biohazardous material if cut and may cause a delay in patient results.

Action

LGC Group notified their sole US consignee (distribution partner) of the recall on 08/25/2023 via email. The notice explained the issue and the risk with the request to notify their end users. The end user was advised to take additional care when handling the glass vials due to the potential of breakages upon receipt, while thawing, or while opening the vials.

Distribution

US Nationwide distribution in the state of Illinois.

Quantity

117155 units