FDA Recall Open, Classified

Atellica CH Systems Fructosamine (FRUC)-For in vitro diagnostic use in the quantitative measurement of glycated protein (fructosamine) in human serum or plasma on the ADVIA Chemistry systems Siemens Material Number (SMN): 11097637

Recall: Z-0102-2021 · Initiated September 22, 2020

Recall

Recall Number
Z-0102-2021
Event Number
86504
Firm
Siemens Healthcare Diagnostics, Inc.
FEI Number
2432235
Product Code
JIT
Status
Open, Classified
Root Cause
Device Design
Initiated
September 22, 2020
Posted
October 13, 2020
Address
511 Benedict Ave, Tarrytown, NY, 10591-5005

Description

Atellica CH Systems Fructosamine (FRUC)-For in vitro diagnostic use in the quantitative measurement of glycated protein (fructosamine) in human serum or plasma on the ADVIA Chemistry systems Siemens Material Number (SMN): 11097637

Reason

Fructosamine QC and patient samples may exhibit a 60 umol/L positive bias across the measuring range compared to predicate assay reference leading to a potential adjustment of hyperglycemic therapy. Positively biased fructosamine results may prompt changes in glucose management that may affect glycemia. In extreme situations, more aggressive management of glycemia may occur and contribute to hypoglycemia.

Action

Siemens Urgent Medical Device Recall (ACHC20-18.A.US.CHC) distributed to customers on 9/22/20, in the United States and an Urgent Field Safety Notice (ACHC20-18.A.OUS.CHC) was distributed to customers outside the United States who have received ADVIA Chemistry Fructosamine in-date reagent lots. For Atellica, an Urgent Medical Device Recall (ACHC20-18.A.US) was distributed to l customers in the United States and an Urgent Field Safety Notice (ACHC20-1.OUS) was distributed to all customers outside the United States who have received Atellica CH in date reagent lots. These letters instruct the customers to: Perform the following actions on the ADVIA Chemistry System 1. Enter the Real-time correction factor in the Analytical Parameters (Chemistry) window. 2. Once the Real-time correction factor has been configured, update QC ranges as described above in the FRUC QC Definition following your laboratory procedures. 3. Update the reference range following your laboratory procedures. 4. Perform a FRUC calibration and process QC. 5. Perform a system back-up. Complete and return the Field Correction Effectiveness Check attached to this letter within 30 days. " Review this letter with your Medical Director. " If you have received any complaints of illness or adverse events associated with the products listed in Table 1, immediately contact your local Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative

Distribution

CA, DE, MD, NC, NJ, NY, PA Foreign: Argentina¿¿¿¿¿¿ Australia¿¿¿¿¿¿ Brazil¿¿¿¿ Canada¿¿¿¿¿ Czech¿Republic¿ Denmark¿¿¿¿¿¿¿¿ Germany¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ Netherlands¿¿¿¿ P.R.¿China¿¿¿¿¿ Paraguay¿¿¿¿¿¿¿ Slovakia¿¿¿¿¿¿¿ Turkey¿¿¿¿¿¿¿¿¿ United¿Kingdom¿ Vietnam¿¿¿¿¿¿¿¿

Quantity

US: 28; OUS: 255 units