cobas p 501 Post-Analytical Unit calculator/data processing module, for clinical use Product Usage: The MODULAR PRE-ANALYTICS system is a fully automated system for sample pre-processing, including centrifuging, aliquoting, and barcode labeling, prior to analysis by MODULAR SYSTEMS analyzers.
Recall
- Recall Number
- Z-0102-2017
- Event Number
- 75186
- Firm
- Roche Diagnostics Operations, Inc.
- FEI Number
- 1823260
- Product Code
- JQP
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- July 15, 2016
- Terminated
- June 5, 2017
- Address
- 9115 Hague Rd, Indianapolis, IN, 46256-1025
Description
cobas p 501 Post-Analytical Unit calculator/data processing module, for clinical use Product Usage: The MODULAR PRE-ANALYTICS system is a fully automated system for sample pre-processing, including centrifuging, aliquoting, and barcode labeling, prior to analysis by MODULAR SYSTEMS analyzers.
Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.
On 7/15/2016, URGENT MEDICAL DEVICE CORRECTION notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, an updated cleaning procedure, and instructions for responding to the formal recall notification. Please contact the Roche Support Network Customer Support Center at 1-800-428-2336 if you have questions about the information contained in this UMDC. On 10/5/2016, URGENT MEDICAL DEVICE CORRECTION UPDATE notifications were sent to the affected customers via courier. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. The notification updated the 7/15/16 notification to include additional affected systems and to provide updated cleaning instructions for all systems via the enclosed Operators Manual addendum.
US Nationwide Distribution
9014 in total