Accumetrics part number 29001 VerifyNow Instrument, printe power supply

Recall: Z-0102-06 · Initiated September 13, 2005

Recall

Recall Number
Z-0102-06
Event Number
33691
Firm
Accumetrics Inc
FEI Number
3001376288
Product Code
JOZ
Status
Terminated
Root Cause
Other
Initiated
September 13, 2005
Posted
October 28, 2005
Terminated
March 3, 2007
Address
3985 B Sorrento Valley Blvd, San Diego, CA, 92121

Description

Accumetrics part number 29001 VerifyNow Instrument, printe power supply

Reason

In 2001, separate reports of mild shock, sparking and smoke from Accumetrics PN 29001 power supply.

Action

Letters and replacement power supplies were sent to consignees by Fedex, 2-day with proof of delivery. End-user customers were instructed to return the affected power supply to Accumetrics, using the provided pre-paid shipping carton. Consignees who do not return the power supply within two weeks of the notification will be contacted by phone to assess compliance with the recall and to determine the status of the product at the site.

Distribution

CA, FL, IL, LA, NH, NV, NY, OK, WA

Quantity

11