FDA Recall Terminated

Plum 360 Infusion System, List number 30010.

Recall: Z-0101-2018 · Initiated October 30, 2017

Recall

Recall Number
Z-0101-2018
Event Number
78312
Firm
ICU Medical Inc
FEI Number
3013319212
Product Code
FRN
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
October 30, 2017
Terminated
August 28, 2019
Address
600 N Field Dr, Lake Forest, IL, 60045-4835

Description

Plum 360 Infusion System, List number 30010.

Reason

(1) Under certain conditions, if a malfunction alarm occurs while the pump is in the "Paused" state waiting for the distal pressure to decrease, the pump cannot be turned off and delivery cannot be restarted until the battery is discharged or is disconnected; and (2) when the user accesses the Preventive Maintenance Screen in Service/Biomed Mode with a Total Delivery Time >1,500 hours, the user will not be able to interact with the device and the device must be power cycled.

Action

The recalling firm issued letters dated 10/30/2017 via UPS on 10/30/2017 notifying their customers of the two issues.

Distribution

Distribution was nationwide to medical facilities. Foreign distribution was made to Canada and Australia. Government and military distribution was also made.

Quantity

21,461 devices