FDA Recall Terminated

Lifescan OneTouch SureStep Meter Blood Glucose monitoring system,Product is distributed by LifeScan, Inc., 1000 Gibraltar Dr., Milpitas, CA 95037

Recall: Z-0101-06 · Initiated September 9, 2005

Recall

Recall Number
Z-0101-06
Event Number
33121
Firm
Lifescan Inc
FEI Number
2939301
Product Code
CGA
Status
Terminated
Root Cause
Other
Initiated
September 9, 2005
Posted
October 28, 2005
Terminated
December 18, 2006
Address
1000 Gibraltar Dr, Milpitas, CA, 95035-6301

Description

Lifescan OneTouch SureStep Meter Blood Glucose monitoring system,Product is distributed by LifeScan, Inc., 1000 Gibraltar Dr., Milpitas, CA 95037

Reason

The LCD of the OneTouch SureStep Meter may display incorrectly, resulting in missing segments on the display. This display failure may result in patients misreading the blood glucose result, or cause a delay in test interpretation, and could cause delays or mistreatment.

Action

On 9/9/05, ***and 4/14/06 (expanded)***the firm notified its consumers, Health Care Professionals (HCPs), pharmacies, and direct accounts (including wholesalers and distributors) by letter and phone calls.

Distribution

Total amount of product distributed in the U.S. is 19,177 units, to distribution centers to be distributed to retailers, Health Care Professionals, wholesalers, and mail order companies. The product was distributed throughout the United States; 29,377 were distributed to Asia, Europe and Latin America.

Quantity

19,177 meters were distributed in the U.S., with 29,377 distributed abroad, to Asia, Europe and Latin America.