CRN Crono Syringe CRN20 with 19G x 1 "Steel Needle"
Recall
- Recall Number
- Z-0099-2015
- Event Number
- 66541
- Firm
- Intra Pump Infusion Systems
- FEI Number
- 3004839315
- Product Code
- FRN
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- October 18, 2013
- Posted
- October 21, 2014
- Terminated
- February 1, 2016
- Address
- 920 Minters Chapel Rd, Ste 200, Grapevine, TX, 76051-1302
Description
CRN Crono Syringe CRN20 with 19G x 1 "Steel Needle"
Product has the potential to have a clear debris on the tip of the needle.
Cane S.p.A. Medical Technology sent an Urgent: Correction and Removal Action Required letter dated October 21, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Actions to be taken: a. Stop using the 19G x 11/2" 1.10 x 40mm steelneedle included in the CRONO Syringe blister pack. B. Safely discard the 19G x 11/2" 1.10 x 40mm steel needle into a Sharps container. c. Use a TRANSFER DEVICE (e.g.,Needleless adapter with female-to-female luer connector) or alternative device to transfer medications into the CRONO syringe. d.For intravenous infusions,we are recommending that a filter also be used during infusions. e. Complete the required 'Acknowledgment and Receipt Form'and fax back to our Distributor;Intra Pump Infusion Systems at 630-845-2768. Contact Information for our US Distributor: INTRA PUMP INFUSION SYSTEMS " Monday through Friday 8:00am to 5:00pm CST,TollFree:1-866- 211-7867 " Direct:630-845-7867 " Fax:630-845-2768 www.lntrapump.com
Worldwide Distribution: US (nationwide) in states of CA, PA, MA, IL, OH and NY; and Internationally to Canada.
25,440 pieces