FDA Recall Terminated

SoftWeb version 4.0.4.10.1. SCC Soft Computer, 5400 Tech Data Drive, Clearwater, FL 33760. Manufacture Date: 05/16/2006.

Recall: Z-0099-2011 · Initiated June 8, 2006

Recall

Recall Number
Z-0099-2011
Event Number
56282
Firm
SCC Soft Computer
FEI Number
1000307651
Product Code
LNX
Status
Terminated
Root Cause
Software design
Initiated
June 8, 2006
Posted
October 20, 2010
Terminated
November 3, 2010
Address
5400 Tech Data Drive, Clearwater, FL, 33760

Description

SoftWeb version 4.0.4.10.1. SCC Soft Computer, 5400 Tech Data Drive, Clearwater, FL 33760. Manufacture Date: 05/16/2006.

Reason

If a case number is re-used in SoftWeb with SoftPath, the SoftPath case is linked to the original MRN in SoftWeb. This causes SoftPath cases being linked to the wrong patient in SoftWeb. Incorrect results on the patient with original MRN and no results on the patient with the second MRN (and the reused case number) are displayed in SoftWeb Results Query. SCC Soft Computer issued a Correction for

Action

SCC Soft Computer issued a Risk-to-Health notification with Client Notification Verbiage to the affected client using SCC Soft Computer's proprietary Task Management System on June 8, 2006. SCC recommended SoftPath/SoftWeb users NOT re-use SoftPath case numbers when cases are cancelled. The generic option setting "Specimen Registration - Standard Settings - Minimum Time to Reactivate Case Number" should be set to "0" so the user is not prompted to reuse the same case number. Client was asked to acknowledge receipt of the task and grant permission to load the required software correction.

Distribution

Nationwide Distribution: 1 hospital in Alaska.

Quantity

1