FDA Recall Terminated

Clea Stand Ceiling used with the Integris X-Ray System H5000, Allura 12/15, Allura 9C

Recall: Z-0099-04 · Initiated October 10, 2003

Recall

Recall Number
Z-0099-04
Event Number
27485
Firm
Philips Medical Systems Sales & Service Region No. America
FEI Number
1000524572
Product Code
EHD
Status
Terminated
Root Cause
Other
Initiated
October 10, 2003
Posted
July 20, 2004
Terminated
January 13, 2005
Address
22100 Bothell Everett Highway, Bothell, WA, 98041

Description

Clea Stand Ceiling used with the Integris X-Ray System H5000, Allura 12/15, Allura 9C

Reason

The cable carrier and cables may fall after repeat collisions with OR lamps, radiation shields or other supported accessories.

Action

The firm sent a letter dated, October 10, 2003 indicating the firm''s service representatives will upgrade the cable carrier by mounting a safety catch inside the cable carrier.

Distribution

The firm distributes to hospitals and medical centers nationwide.

Quantity

536 units