Clea Stand Ceiling used with the Integris X-Ray System H5000, Allura 12/15, Allura 9C
Recall: Z-0099-04
·
Initiated October 10, 2003
Recall
- Recall Number
- Z-0099-04
- Event Number
- 27485
- Firm
- Philips Medical Systems Sales & Service Region No. America
- FEI Number
- 1000524572
- Product Code
- EHD
- Status
- Terminated
- Root Cause
- Other
- Initiated
- October 10, 2003
- Posted
- July 20, 2004
- Terminated
- January 13, 2005
- Address
- 22100 Bothell Everett Highway, Bothell, WA, 98041
Description
Clea Stand Ceiling used with the Integris X-Ray System H5000, Allura 12/15, Allura 9C
Reason
The cable carrier and cables may fall after repeat collisions with OR lamps, radiation shields or other supported accessories.
Action
The firm sent a letter dated, October 10, 2003 indicating the firm''s service representatives will upgrade the cable carrier by mounting a safety catch inside the cable carrier.
Distribution
The firm distributes to hospitals and medical centers nationwide.
Quantity
536 units